Hioki battery tester BT3563A | APC Technology Group

BT3563A

BT3563A

Hioki

Battery Hitester BT3563A - Fully Automated Production Line Testing of Large Packs for xEVs or Large Packs of up to 300 V

This item is in stock and can be delivered in 3-5 business days

Contact our team for details on [email protected]

Contact us for lead time

Contact our team for stock availability and product options.

Price £2,924.00
More Information
Attribute Value
Key Features
  • Simultaneous measurement of internal resistance and open circuit voltage.
  • Fully automated production line testing of large packs for xEVs or large packs of up to 300 V.
  • Resistance measurement ranges: 3 mΩ/30 mΩ/300 mΩ/3 Ω/30 Ω/300 Ω/3 kΩ
  • Voltage measurement ranges: 6 V/60 V/300 V
  • Equipped with LAN
Notes Measurement leads are not included. Purchase the appropriate lead option for your application separately. The male (system side) of the EXT I/O connector is also available. Please contact APC Technology Group.
Basic Specifications
Resistance measurement ranges 3 mΩ(Max. display: 3.1000 mΩ, resolution: 0.1 μΩ, measurement current: 100 mA)
30 mΩ(Max. display: 31.000 mΩ, resolution: 1 μΩ, measurement current: 100 mA)
300 mΩ(Max. display: 310.00 mΩ, resolution: 10 μΩ, measurement current: 10 mA)
3 Ω(Max. display: 3.1000 Ω, resolution: 100 μΩ, measurement current: 1 mA)
30 Ω(Max. display: 31.000 Ω, resolution: 1 mΩ, measurement current: 100 μA)
300 Ω(Max. display: 310.00 Ω, resolution: 10 mΩ, measurement current: 10 μA)
3 kΩ(Max. display: 3.1000 kΩ, resolution: 100 mΩ, measurement current: 10 μA)

Basic accuracy: ±0.5% rdg ±10 dgt(3 mΩ range), ±0.5% rdg ±5 dgt(30 mΩ range or more)
Measurement frequency: 1 kHz ±0.2 Hz
Measurement method: AC four-terminal method
Voltage measurement ranges 6 V(Max. display: 6.00000 V, resolution: 10 μV)
60 V(Max. display: 60.0000 V, resolution: 100 μV)
300 V(Max. display: 300.000 V, resolution: 1 mV)

Basic accuracy: ±0.01% rdg. ±3 dgt.
Response time 10 ms
Sampling period Ω or V (60 Hz): 4 ms (EX.FAST), 12 ms (FAST), 35 ms (MEDIUM), 150 ms (SLOW)
ΩV (60 Hz): 8 ms (EX.FAST), 24 ms (FAST), 70 ms (MEDIUM), 253 ms (SLOW)

Ω or V (50 Hz): 4 ms (EX.FAST), 12 ms (FAST), 42 ms (MEDIUM), 157 ms (SLOW)
ΩV (50 Hz): 8 ms (EX.FAST), 24 ms (FAST), 84 ms (MEDIUM), 259 ms (SLOW)
Functions Contact check, Zero adjustment (±1000 counts), Pulse measurement, Comparator (Hi/ IN/ Lo), Statistical calculations (Max. 30,000), Delay, Average, Panel saving/loading, Memory storage, LabVIEW® driver
Interfaces LAN (TCP/IP, 10BASE-T/100BASE-TX)
RS-232C (Max. 38.4 kbps, Available as printer I/F)
EXT I/O (37-pin Handler interface)
Analog output (DC 0 V to 3.1 V)
Power supply 100 to 240 V AC, 50 Hz/60 Hz, 35 VA max.
Dimensions and mass 215 mm (8.46 in) W × 80 mm (3.15 in) H × 295 mm (11.61 in) D, 2.4 kg (84.7 oz)
Instrument Type
More Information
Attribute Value
Data Sheet Click here to view

Related Products

  1. L2020
  2. BT3554-51
  3. RM3548
  4. 9465-90
  5. BT3554-52