5012D
5012D
Bird
150mW - 150W Avg, 400W Peak Wideband RF Power Sensor
Peak Wideband RF Power Sensor 150mW - 150W Avg, 400W
The 5012D RF Power Sensor never requires field calibration, only requires calibration once per year and is traceable to the National Institute of Standards and Technology (NIST). The WPS measures True Average Power, Peak Power, and Duty Cycle directly with exceptional accuracy and uses these precise measurements to calculate a wide range of other important factors, such as VSWR, Return Loss, Reflection Coefficient, Crest Factor, Average Burst Power, and CCDF.
Included Accessories:
- Bird RF 5012D Manual
- VPM3 - Virtual Power Meter Software and Manual
- 10 ft. USB Cable
Please note:
The asterisk (*) displayed in the Product Specifications below indicates (for temperatures above 35°C or below 15°C add 3.0% to stated accuracies) - this includes the entire "Accuracy-Peak Envelope Power" spec as well.
Click below to request a quote for this item. Or for product advice, stock and lead time enquiries call our team on 0330 313 3220.
Attribute | Value |
---|---|
Manufacturer | Bird |
Certificates | EMC Directive (2004/108/EC) European Standard: EN 61326— Electrical Equipment for measurement, control and laboratory use; EMC Requirements Test Spec (for radiated immunity): EN 61000-4-3—Testing and measurement techniques - 10V/meter CE Mark RoHS |
Impedance Ohm | 50.0000 |
Accuracy - Peak Envelope Power | burst width > 200 μs: ± 7% of reading, + 0.20 W 1 μs < burst width < 200 μs: ± 10% of reading, + 0.40 W 0.5 μs < burst width < 1 μs: ± 15% of reading, + 0.40 W burst width < 0.5 μs: ± 20% of reading, + 0.40 W |
Average Power | Average Forward Power Range: 150 mW - 150 Watts Avg, 400 Watts Peak *Accuracy, Average Forward Power: ± 4% of reading, + 0.05 W Minimum Forward Power for Reflected Measurement: 0.5W Return Loss: 0.0 to 23 dB VSWR: 1.15 to 99.9 |
Burst Average Power | Burst Average Forward Power Range: 4 W - 150 Watts Avg Burst Width: 1 μs to 5 ms Repetition Rate: 5 Hz, Min Duty Cycle (D): .002 to 1.0 *Accuracy, Burst Average Power: ± 6% of reading, + 0.05 Watts |
Complementary cumulative distribution function | CCDF Measurement Range: 0.1 to 100% Threshold Measurement Range: 4.0 - 400 W Measurement Uncertainty: ± 2% *Level Set Accuracy: As Peak Envelope Power Accuracy + 2.0% |
Crest factor | Crest Factor Measurement Range: 150mW - 150 Watts *Accuracy, Crest Factor: Linear Sum of Peak and Average Power Accuracies |
Directivity (dB) | 30 dB up to 3.0 GHz 28 dB from 3.0 to 4.0 GHz |
Insertion Loss | <0.05 dB from 0.35 to 1.0 GHz <0.1 dB from 1 to 4 GHz |
Insertion VSWR | <1.05 from 0.35 to 2.5 GHz <1.10 from 2.5 to 4 GHz |
Peak envelope power range | 4.0 - 400 W |
Power range | 150 mW to 150 Watts Avg., 400 Watts Peak |
Connectors | N Female (Both) |
Temperature Range | Operating -10 °C to +50 °C (+14 °F to +122 °F) Storage -40 °C to +80 °C (-40 °F to +176 °F) |
Power Supply | USB Port: less than one low-power USB load DC Input Connector: 7-18 VDC at less than 0.1A |
Dimensions | 4.8 in x 4.6 in x 1.3 in (122 mm x 117 mm x 33 mm) |
Attribute | Value |
---|---|
Data Sheet | Click here to view |