ELSE NUCLEAR VEBOS | APC Technology Group

VEBOS

VEBOS

ELSE NUCLEAR

Whole Body In-Vivo Counting System

The VEBOS system is designed to perform a complete scan of individuals standing in front of a vertically moving detection equipment.

The measuring equipment is composed of 2 liquid-nitrogen-cooled HPGe detectors, each coupled to a 30 l nitrogen dewar, and managed by a multi-channel analyser.

The detectors are shielded with lead collimators in order to reduce the contribution due to environmental background. The collimators can also be used to define a detection field of view, according to the measurement requirements (e.g. full body scan, hot spot deep analysis, etc.).

The detectors and related components are lodged on a mechanically moving shelf, activated by the system software according to the pre-set measurement sequence.

The mechanical structure of VEBOS includes a main frame which acts both as supporting structure as well as shielding from external background contribution.

The primary user interface is provided by a built-in panel PC with proprietary software allowing the complete management of the system, i.e. activation and control of the measurements, display/archive of the results, etc.

Several safety devices are implemented in VEBOS: a base platform that detects an individual presence, interlocks at the maintenance door, and buttons on the holding handles to enable the movement of the detectors.

Key Features

  • Measure of total body activity and committed effective dose
  • Fully automatic scanning system
  • Alternative detection equipment options
  • Detectors: two liquid-nitrogen-cooled HPGe
  • Low-background lead shielding/collimators
  • Extensive safety devices
  • Several operating modes available according to required activities
  • Alarm column
  • Handles for helping the person to keep a suitable position during measurements
  • Door for easy access to the detectors for maintenance activities
  • Eyebolts for easy deployment of the system
Contact us for lead time

Click below to request a quote for this item. Or for product advice, stock and lead time enquiries call our team on 0330 313 3220.

More Information
Attribute Value
Energy Range 40 KeV - 10 MeV
Dimensions 2000 x 2720 x 1800mm (W x H x D)
Options

Alternative detection equipment options, e.g.: liquid-nitrogen-cooled HPGe detectors, NaI(Tl) scintillators, etc.

Accessories BOMAB phantom for calibration
Detection Unit

Detector type: electro-cooled coaxial P-type HPGe detectors or NaI(Tl) scintillators

Number of units: 2

Lead shielding/collimator thickness: 50 mm

Relative efficiency: ≥ 25%

Energy resolution (122 keV): 0.82 keV

Energy resolution (1.33 MeV): 1.85 keV

Peak to Compton: 56:1

Detectors endcap size: 70 mm

More Information
Attribute Value
Data Sheet Click here to view