ELSE NUCLEAR LEM | APC Technology Group

LEM

LEM

ELSE NUCLEAR

Liquid Effluent Monitoring System

The LEM system allows to sample the liquid effluents and to perform a gross gamma measurement of the specific activity in Marinelli geometry.

LEM system is composed of the following main parts:
Stainless steel frame
Electrical and command board with touch-screen panel PC
NaI(Tl) detector, 1 l Marinelli, 5 cm thick lead shielding well
Self-priming pump
Software for system management, data acquisition and processing
The measurements are visualised in real time by the software, in terms of specific/total activity or count rate/integrated counts.

LEM status and parameters are managed by the ELSE NUCLEAR software.

The system provides also I/O contacts through dedicated connectors:
Good functioning status output
Alarm status output
Pump activation input from customer PLC
Spare available I/O contacts (to be defined when necessary)
The software provides a calibration routine, to be used with a Marinelli calibration source (available as accessory).

A test program is also available, separate from the main application, to be used for maintenance or periodical quality controls.

Key features

  • Fully customizable sampling cycle
  • Simple and reliable effluent monitor
  • I/O contacts for status and commands
  • 1 liter Marinelli beaker measurement geometry
  • 2”x2” Nal(Tl) scintillator
  • Lead shielding well
  • Panel PC and relay contacts for I/O signals
  • Efficiency calibration function for reference radionuclide
Contact us for Lead Time

Click below to request a quote for this item. Or for product advice, stock and lead time enquiries call our team on 0330 313 3220.

More Information
Attribute Value
Technical Specifications
  • Detector type: 2”x2” Nal(Tl) scintillator
  • Cylindrical lead well, 5 cm thickness, composed by several rings for easy assembling and installation
  • Plexiglass, 1 liter Marinelli beaker, fittings included
  • Electronics for HV and signal processing
  • Electric safety, isolation and waterproofing
  • Protection grade: IP67 (underwater parts), IP44 (non-underwater parts)
Dimensions

Stainless steel frame:
WxDxH = 80 x 70 x 150 cm
Lead well:
Ø x H = 32 x 30 cm (ext); 22 x 20 cm (int)

Weight

Stainless steel frame:
max weight = 500 kg
Lead well:
max weight = 300 kg

Options

3 liters Marinelli beaker
10 cm lead shielding for further lowering of the MDA
MCA as an alternative to SCA

Accessories

Calibration source (isotopes and activity to be defined):
gel matrix in Marinelli beaker
Cs-137 point source, < 10 kBq, for periodical quality
controls

Warranty Warranty extension from 12 months to 24 months_x000D_
Measurement Features

Measurement unit: specific activity (Bq/l) or count rate (cps or cpm)
Gross gamma measurement of the specific activity
Efficiency in measurement geometry: >1.5% (Cs-137)
Response in measurement geometry: >12 cps/kBq (Cs-137)
Energy range: 30 keV – 2 MeV
Measurement range: 3.7 x 10 – 3.7 x 10 Bq/l
“Fail safe”: alarm and interlock activation in case of power supply
interruption
Detector can easily be disassembled for maintenance activities

Dimensions

Stainless steel frame:
WxDxH = 80 x 70 x 150 cm
Lead well:
Ø x H = 32 x 30 cm (ext); 22 x 20 cm (int)

Weight

Stainless steel frame:
max weight = 500 kg
Lead well:
max weight = 300 kg

More Information
Attribute Value
Data Sheet Click here to view

Related Products

  1. WDMS NT-VK
  2. NAUSICAA 2IC