The Hioki DT4200 Digital Multimeter Series consist of a complete lineup of 9 models ranging from professional to industrial to pocket meters, delivering a super fast response rate and safety features that take electrical testing to a higher level. The DT4252 is a digital multimeter designed specifically for general purpose testing, with ±0.3% DC V accuracy, direct current input and wide 40 Hz to 1 kHz bandwidth. CAT IV 600 V, CAT III 1000 V
Contact us for lead time

Click below to request a quote for this item. Or for product advice, stock and lead time enquiries call our team on 0330 313 3220.

More Information
Attribute Value
Key Features
  • ±0.3% DC V basic accuracy, wide 40 Hz to 1 kHz AC V frequency characteristics
  • Measure up to 10A with direct input
  • Dual display lets you check voltage and frequency simultaneously
  • Low-pass filter cuts high harmonics (when measuring inverter fundamental waveforms)
  • USB communications function supports PC measurements (optional)
Basic Specifications
DC Voltage range 600.0 mV to 1000 V, 5 ranges
Basic accuracy: ±0.3 % rdg. ±5 dgt.
AC Voltage range 6.000 V to 1000 V, 4 ranges, Frequency characteristics: 40 Hz - 1 kHz
Basic accuracy 40 - 500 Hz : ±0.9 % rdg. ±3 dgt. (True RMS, crest factor 3)
AUTO AC/DCV N/A
Resistance range 600.0 Ω to 60.00 MΩ, 6 ranges, Basic accuracy: ±0.7 % rdg. ±5 dgt.
DC Current range 6.000 A / 10.00 A, 2 ranges, Basic accuracy: ±0.9 % rdg. ±5 dgt.
AC Current range 6.000 A / 10.00 A, 2 ranges,
Basic accuracy 40 - 500 Hz : ±1.4 % rdg. ±3 dgt.
(True RMS, crest factor 3, 40 Hz to 1 kHz)
AC Current range (use with Clamp on probes) N/A
Voltage detection (50/60 Hz) N/A
Capacitance range 1.000 μF to 10.00 mF, 5 ranges, Basic accuracy: ±1.9 % rdg. ±5 dgt.
Frequency range AC V, AC A measurement: 99.99 Hz (5 Hz or more) to 9.999 kHz, 3 ranges, 99.99 kHz (only for AC V)
Basic accuracy: ±0.1 % rdg. ±1 dgt.
Continuity check Continuity threshold [ON]: 25 Ω or less (Indicate buzzer sound, red LED), [OFF]: 245 Ω or more
Response time: 0.5 ms or more
Diode test Open terminal voltage: 5.0 V or less, Testing current 0.5 mA or less,
Threshold of forward voltage: 0.15 V to 1.5 V
Other functions Filter function, display value hold, auto hold, MAX/MIN/Average value display, relative value display, auto-power save, USB communication (option)
Display Main and sub displays: 4-digits LCD, max. 6000 digits, bar graph
Display refresh rates 5 times/s (Capacitance measurement: 0.05 to 5 times/s, depending on measured value, Frequency: 1 to 2 time/s, Temperature: 1 time/s )
Power supply LR03 alkaline batteries ×4, Continuous use: 130 hours (back light OFF)
Dimensions and mass 84 mm (3.31 in)W × 174 mm (6.85 in)H× 52 mm (2.05 in)D, 390 g (13.8 oz) (with batteries and holster)
Accessories Test lead L9207-10 ×1, Holster ×1, Instruction manual ×1, LR03 alkaline battery ×4
More Information