Tektronix TIVP02L IsoVu 200MHz Isolated Probe 10m|APC Technology Group

TIVP02L

TIVP02L

Tektronix

Tektronix IsoVu 200 MHz Isolated Probe with 10 m cable

Tektronix IsoVu probe technology delivers accurate differential measurements up to ±2500 V on reference voltages slewing ±60 kV at 100 V/ns or faster. With the IsoVu Generation 2 design, you get all the benefits of IsoVu technology at 1/5 of the size of first-generation probes.

With versatile MMCX connectors and an unmatched combination of bandwidth, dynamic range, and common mode rejection, IsoVu Gen 2 probes are setting new standards for isolated probe technology and enabling wide bandgap power designs using SiC and GaN.

High Voltage and High Bandwidth

With traditional differential probes, you had to choose between high bandwidth or high voltage levels. IsoVu probes, with their shielded coaxial cable and isolation, provide high bandwidth and a differential voltage range of ±2500 V. IsoVu Gen 2 offers bandwidths of 200 MHz, 500 MHz, and 1 GHz to fit your budget and performance needs.

High Performance and Convenient Connections

The Tektronix IsoVu probe tips have a range of connections and accessories that offer high performance and accessibility. The probes can connect directly to MMCX connectors, which are inexpensive and widely available. This makes for stable, hands-free test points and offers high bandwidth and common mode rejection. The solid metal body shields the center conductor and minimises ground loop area for the lowest interference possible.

Other accessories are available to adapt the probe tips to a wide range of connections. Additional 0.100" and 0.200" spaced square-pin tips are available for applications that require greater than ±250V differential voltage. When not using a tip, the sensor head has 1 MΩ and 50Ω switchable termination at the probe's SMA connector. This feature effectively adds an isolated channel to any compatible oscilloscope.

Floating Measurements in Power Converter and Motor Drive Designs

Making high-side measurements in half-bridge power converters is challenging because the source or collector to which the measurements are referenced is slewing rapidly up and down. Wide bandgap devices like SiC and GaN FETs are even harder to measure because they can switch high voltages in a few nanoseconds. Noise from this rapidly changing common mode voltage leaks into the differential measurements and hides details on VGS and VDS. IsoVu probes have unmatched common mode rejection at full bandwidth that lets you see signal details, often for the first time.

Contact us for lead time

Click below to request a quote for this item. Or for product advice, stock and lead time enquiries call our team on 0330 313 3220.

More Information
Attribute Value
Product Segment IsoVu Isolated Probes
Product Family IsoVu Isolated Probes
Key Features
  • 160 dB (100 million to 1) CMRR at DC
  • Up to 120 dB (1 million to 1) CMRR at 100 MHz
  • Up to 80 dB (10,000 to 1) CMRR at 1 GHz
  • ±60 kV Common Mode Voltage range
  • Up to ±2500 V differential input Voltage range
  • Up to ±2500 V offset range
Bandwidth 200 MHz
Applications
  • Half/Full bridge designs using SiC or GaN, FETs, or IGBTs
  • Floating measurements
  • Power converter design
  • Power device evaluation
  • Switching power supply design
  • Inverter design
  • Motor drive design
  • Electronic ballast design
  • EMI and ESD troubleshooting
  • Current shunt measurements
Cable Length 10m
Accessory Type Probe
More Information
Attribute Value
Data Sheet Click here to view

Related Products

  1. TRCP0300
  2. TDP7704
  3. TIVP1L
  4. DPO7OE2
  5. TCP0150