ELSE NUCLEAR MISTRAL XM | APC Technology Group

MISTRAL XM

MISTRAL XM

ELSE NUCLEAR

Air Activation Monitoring System

MISTRAL XM is a system designed to sample and monitor the gamma activity concentration resulting from air activation, in Marinelli geometry.

In particular, MISTRAL XM Rooms is designed for free air monitoring, whereas MISTRAL XM Stack is used to monitor the air expelled from a chimney or a stack. In both cases, “X” indicates the number of sampling points: up to 8 (Rooms) or 1 (Stack).
The system is composed of:

acquisition and processing unit: APU
sampling and detection unit: SDU-XM
The APU is the main user interface with the system, and it consists in a panel PC mounted on a wheeled support.


The SDU-XM includes the NaI(Tl) detector with MCA, the system electronics, a lead shielding well, a pump for air sampling with a control flow meter, and an array of electrovalves (if needed). The system also manages the expulsion of the monitored air.
The accessory flow rate meter STACK-DFM can be included in the stack version, to calculate the specific activity of the expelled air volume.

The software installed on the PC displays in real time the measurement, controls the system status, and allows to set the operative parameters, such as the alarm thresholds. The user can define specific regions of interest (ROIs) and thus determine the specific activity (Bq/g) of the sampled air.

Main features:

  • Extremely high sensitivity
  • Rooms and stack monitoring
  • Fully configurable work cycle
  • Lead well for 3 l Marinelli geometry
  • Detector: NaI(Tl) 2”x2” scintillator with MCA
  • Electrovalves for automatic sampling of up to 8 rooms 
  • Acquisition/processing SW, display of energy spectra
  • Relays for external devices
  • MDA (F-18): < 1 Bq/l (300 s) 
Contact us for lead time

Click below to request a quote for this item. Or for product advice, stock and lead time enquiries call our team on 0330 313 3220.

More Information
Attribute Value
Energy Range

Energy range: 30 keV ÷ 2 MeV

ICPT Gamma Detector

Sampling and detection unit

  • Detector: 2”x2” NaI(Tl)
  • MDA (F-18): < 1 Bq/l in 5 min measurement
  • Energy range: 30 keV ÷ 2 MeV
  • Temperature range: 0 - 40 °C
  • Geometry: 3 liters Marinelli beaker
  • Shielding: lead well, 100 mm thick
Dimensions

Overall dimensions (WxHxD): 100 x 120 x 60 cm

Options
  • LaBr (Ce ) scintillator instead of NaI(Tl), for higher sensitivity and better gamma spectrum resolution
  • Alternative version with remote APU (wheeled support for 19" PC, 60 x 170 x 60 cm, approx. 80 kg)
Accessories

Stack flow rate meter for released activity calculation: STACK-DFM

More Information
Attribute Value
Data Sheet Click here to view