9272-05
9272-05
Hioki
Clamp on Sensor
Ideal for Measuring AC Current with Low Frequencies such as Inverter Control Devices
Hioki current sensors are best-in-class devices for use with power meters, Memory HiCorders and high performance oscilloscopes. The 9272-10/9272-05 are a 1 Hz to 100 kHz AC current sensor ideal for measuring low frequency equipment such as inverters.
CAT III 600 V
Use with an instrument with a high input impedance of 1 MΩ or higher.
Key Features
- Superior low frequency and phase characteristics suitable for testing the current and power of inverter control devices
- Wide 1 Hz to 100 kHz frequency bandwidth perfect for harmonic analysis, FFT analysis and waveform monitoring (AC only)
Contact us for lead time
Click below to request a quote for this item. Or for product advice, stock and lead time enquiries call our team on 0330 313 3220.
Attribute | Value |
---|---|
Power Supply | Power suppled via the Power Analyzer PW8001, PW6001, PW3390, Sensor Unit CT9555, CT9556, CT9557, or 3CH CURRENT UNIT U8977 |
Included Accessories | Carrying case 9355 ×1, Instruction manual ×1, Mark bands ×6 |
Max Allowable Input | 50 A rms (at 20 A range), 300 A rms (at 200 A range) |
Dimensions and Weight | 78 mm (3.07 in) W × 188 mm (7.40 in) H × 35 mm (1.38 in) D, 430 g (15.2 oz), cord length: 3 m (9.84 ft) |
Basic Accuracy | Amplitude: ±0.3 % rdg. ±0.01 % f.s. Phase: ±0.2 ° (45 to 66 Hz) |
Power Consumption | 5 VA Max. (when measuring 200 A) |
Rated Current | 20 A AC, or 200 A AC (selectable) |
Frequency Bandwidth | 1 Hz (±2 % rdg. ±0.1 % f.s.) to 100 kHz (±30 % rdg. ±0.1 % f.s.) |
Output Voltage Rate | 2 V/20 A or 2 V/200 A (This device outputs AC voltage via the Sensor Unit) |
Core Diameter | φ 46 mm (1.81 in) |
Attribute | Value |
---|---|
Data Sheet | Click here to view |