Bird RF 7022: 350 MHz to 6 GHz, Statistical RF Power Sensor | APC Tech

7022

7022

Bird RF

350 MHz to 6 GHz, Statistical RF Power Sensor

350 MHz to 6 GHz, Statistical RF Power Sensor

Measure and Maintain Complex Wireless Systems

The 7022, Statistical RF Power Sensor is a rugged, easy-to-use field instrument that uses statistical sampling techniques. In modern communications, signal average power, although important is no longer a sufficient control variable.

These communication signals often require more diagnostic measurements. In the statistical mode, the power sensor measures the percentage of time the signal exists at a specific peak-to-average ratio. In the time domain mode, the sensor adds a variety of functions similar to an oscilloscope to our standard suite of measurements

The statistical power sensor can be used in all known communication formats and accurately measures a wide range of RF measurements for non-periodic signals such as EVDO, UMTS, LTE, and HDTV. Like all Bird Sensors, its calibration is also traceable to NIST standards with no field calibration required.

Modes:

  • Time Domain Mode
  • Statistical Mode
  • Average Power Mode

(See Datasheet for more details)


Applications:

Analog Cellular, Digital Cellular, 3G, 4G, Tetra, APCO/P25 Phase 1 & 2, DMR, MOTOTRBO, Tunking, CDMA, TDMA, WCDMA, GSM, Transportation, Tactical Military, Radar, Avionics, Marine, LMR, Analog Broadcast, Digital Broadcast, GSM, GPRS, EDGE, UMTS, HSDPA, Bluetooth, Fire, GPS, NPSPAC, Paging, Public Safety, Telematics, Utilities, WIMAX, WLAN, EVDO, UMTS, LTE, and HDTV.

Standard Accessories:

  • Bird RF 7022 Manual
  • VPM3 Virtual Power Meter Software and Manual
  • SeaLatch USB Cable and BNC Trigger Cable
Contact us for lead time

Click below to request a quote for this item. Or for product advice, stock and lead time enquiries call our team on 0330 313 3220.

More Information
Attribute Value
Manufacturer Bird
Impedance Ohm 50.0000
Directivity <-30 dB, 350 to 1000 MHz (min) <-28 dB, 1000 to 6000 MHz (min)
Frequency Measurement Accuracy ± 3MHz with CW signals
Frequency Range 350 MHz to 6 GHz
Insertion Loss 0.05 dB (max)
Insertion VSWR 1.065 350 to 2000 MHz (max) 1.12, 2000 to 6000 MHz (max)
Peak average ratio 12 dB, absolute peak power limited to 1500W
Power range 0.25 W to 500 W average, Average Power Rating limited (see datasheet)
Connectors Type-N (Female)
Data logging Yes, with the VPM3, Virtual Power Meter Software
Dynamic Range 33 dB
Temperature Range Operating: –10 to +50 °C (+14 to +122 °F) Storage: –40 to +80 °C (–40 to +176 °F)
Calibration Factory: NIST Traceable Field: No Field Calibration Required
Dimensions 5.9 in x 4.8 in x 1.3 in (150 mm x 122 mm x 33 mm)
More Information
Attribute Value
Data Sheet Click here to view

Related Products

  1. 5019D
  2. 5018D
  3. 7020-1-030301
  4. 5014
  5. 5012D